To access the Equipment Info program, select the equipment name in the working set and then click on "Eqp Info" in the working set menu.
Property rgmode=3 (software regulation via NMR probes):
The mass control window is used in the same manner as for the GPS. After each mass change, the field may take 5-10 minutes to stabilise. After each field change the magnet field is cycled (message cycling magnet appears in the mass control window). During cycling the power supply is operating under current regulation: the field is changing too rapidly for the NMR to lock, so there is no field readout. The cycling takes 2-3 minutes, depending on the field precision required and the setup of the software. When field regulation resumes, AQN returns the NMR readout. After a few minutes more, the desired field is reached.
As always, close the beam-gate during mass changes.
Note that the magnets take 5-10 minutes to stabilise after each field change. It is possible to change the field in both magnets simultaneously (ie. there is no need to wait for MAG90 to stabilise changing MAG60). It is also possible to change the requested field (CCV) before the magnet has stabilised -- the magnet cycling routine will be automatically interrupted and restarted.
Re-sending a field request without changing the CCV will be ignored. To provoke a magnet cycle without changing the field, open the field control knob (HRS.MAG90 and HRS.MAG60 in the HRS working set), then request a field change of +0.00001 T, immediately followed by -0.00001 T.
The field range of the NMR probes is 0.09 - 0.52 T, corresponding to a mass range of A = 7 - 217 at 60kV (A = 13 - 432(!) at 30kV).
The mass-control program always shows the field error, which is just the difference between CCV and AQN. Once stable, the field-error should be less than 10 uT and the beamgate may be opened.
Whenever the User Beamgate is open, slit SL470 should be IN (access via the WG470 control window).
Once MAG90 has stabilised, calibrate MAG60 on the centre of SC680. Ensure that the Mass and High-voltage properties are the same as those used for MAG90.
Note the High-Voltage and Mass-Factor values for each magnet in the logbook!
In case the magnet calibration has been totally lost, click "Set Default Values" in the mass control menu. Then fine-tune the calibration using the procedure above.
Using the HRS working set, check the CCV of the power supply of the affected magnet, YHRS.SEPMAG90DEG or YHRS.SEPMAG60DEG. If it reads 470 or above, reduce it by 100A.
Send a new field request and check that the magnet is cycling (see above).
The control software in the DSC has jammed. Either reload the DSC, or get a controls expert to restart the hrsmagrt process. Once restarted, the field CCV will be set to 0.00010 T. Send a new field request and check that the magnet is cycling (see above).
To switch to hardware field regulation:
In BH-15 mode the Mass Control window is used in the same manner as for software regulation. The field stabilises more rapidly, but the calibration is extremely poor. Note that when calculating mass factors, the field AQN from the NMR probes is used (whereas the CCV is used for the GPS). In hardware field regulation the NMR probes are taken out of the feedback loop, and therefore the AQN does not exactly match the CCV. Outside the range 0.09 - 0.52 T the NMR probes will not read at all.